APPLICATIONS
- Acquires very low signals in disturbed areas needing galvanic isolated amplifiers
- Measuring : distortion, pressure, force (gauge sensor)
DESCRIPTION
E300F4 amplifier conditioner can be used twofold:
- strain gauge sensor conditioner
- high precision amplifier for low level voltage
Galvanic isolation gives an excellent common mode rejection.
Gain, filter, sensor supply are set up manually (buttons at the front). Each amplifier has a 220VAC supply.
TECHNICAL FEATURES
Technical features | Low drift | 0.2mV/°C |
| Low noise | |
| Good stability | 5ppm |
| High common mode rejection | |
| Switchable gain | from 1 to 2000, progression 1-2-5… |
| Accuracy | 0.02% |
| Output filter | from 1 to 7KHz (7 positions) |
| Bandwidth | |
| Bridge balance | 0-100KHz |
| Input/output isolation | counter voltage |
| Anti short-circuiting output | ±500V |
| Format | |
| Input features | Camac rack 19’’ 4U with supply sock at the rear |
| Switchable at the front | |
| Stability | |
| Accuracy | |
| Linearity | 0.02% extrapolation by 10-turn potentiometer |
| Gain | 0.01% |
| Drift temperature | |
| According to time | |
| Zero | |
| Linear | |
| Max without damage | |
| Voltage | |
| Max voltage | |
| Common mode | ±500V (input / output) |
| Common mode rejection | |
| Current | Rd=0Ω DC 160dB 50Hz 140dB |
| Input current | |
| Resistance | 1nA max |
| Source resistance | |
| Differential input resistance | |
| Capacity | |
| Capacity | |
| Noise | |
| Noise | |
| Bridge supply | |
| Wiring | |
| Possibility to supply with current | 2 or 4 wires |
Configuration| Voltage supply | |
| Extrapolation of ranges by 10-turn potentiometer from 1 to 2000, positions: 1-2-5… | |
| Current supply | |
| From 0 to 15V DC | |
| Regulation | By switch (1V not available) |
| Regulation line | |
| Regulation charge | 0.005% for ±1% DV |
| Stability | 0.02% for 50 to 100% DV |
| Stability | |
| Drift | ±0.002%/°C |
| Drift | |
| Range | From 0 to ±15mV RTI by switch (1mV not available) |
| Stability | Range extrapolation by 10-turn potentiometer |
| Stability | |
| Output features | ±0.002%/°C |
| Configuration | |
| 2 outputs | |
| Resistance | Filtered output: 2nd order butterworth filter (4th order optional) Non-filtered output |
| Output resistance | |
| Voltage | <0.1Ω |
| Voltage | |
| Current | ±10V mini |
| Current | |
| Capacity | ±5mA |
| Loading capacity | |
| Bandwidth | 0.1mF max |
| Bandwidth | |
| Control command | Non-filtered. From 0 to 100KHz for -3dB |
| 1 logical status | |
| Mechanic features | Input/control switch |
| Weight | |
| Size | 1800g (nominal) |
| Connector | see picture |
| Rack | Female connector FRB 50 contacts |
| Dimensions | |
| Size | |
| Configuration | 19’’, 4U |
| Receives 8 x E300F4 units | |
| Network connection | |
| Connectors at the rear | 220V single-phase, ±10% 50VA |
| Standard | |
| Filtered output | input + bridge: 8-male Jeager connectors (533272) |
| Non-filtered output | 1 x miniature male 27-male Jeager connectors |
| Control relay command / measuring | 1 x miniature male 27-male Jeager connectors |
| Control input | 1x 25-male connectors |
| Power socket | 1x 3-male Jeager connectors |
Other configurations on demand| Environment | |
| Use temperature | |
| Storage temperature | from 0 to +60°C |
| from -55 to +85°C |